Cherenkov Light Emission in Pure Cherenkov Emitters for Prompt Gamma Imaging
Document Type
article
Source
IEEE Transactions on Radiation and Plasma Medical Sciences. 8(1)
Subject
Biomedical and Clinical Sciences
Engineering
Clinical Sciences
Biomedical Engineering
Oncology and Carcinogenesis
Cherenkov emitters
Cherenkov light
prompt-gamma imaging
proton range verification
Prompt-gamma imaging
Proton range verification
Language
Abstract
Proton range verification (PRV) in proton therapy by means of prompt-gamma detection is a promising but challenging approach. High count rates, energies ranging between 1 MeV and 7 MeV, and a strong background complicate the detection of such particles. In this work, the Cherenkov light generated by prompt-gammas in the pure Cherenkov emitters TlBr, TlCl and PbF2 was studied. Cherenkov light in these crystals can provide a very fast timing signal with the potential to achieve very high count rates and to discern between prompt-gammas and background signals. Crystals of 1×1 cm2 and thicknesses of 1 cm, 2 cm, 3 cm and 4 cm were simulated. Different photodetector configurations were studied for 2.3 MeV, 4.4 MeV, and 6.1 MeV prompt-gammas. TlCl achieved the greatest number of detected Cherenkov photons for all energies, detector dimensions, and photodetector efficiency modeling. For the highest prompt-gamma energy simulated, TlCl yielded approximately 250 Cherenkov detected photons, using a hypothetical high-performance photodetector. Results show the crystal blocks of 1 cm × 1 cm × 1 cm have greater prompt-gamma detection efficiency per volume and a comparable average number of detected Cherenkov photons per event.