KOR

e-Article

Towards an Algorithm-Based Tailored Treatment of Acute Neonatal Hyperammonemia
Document Type
article
Source
Toxins, Vol 13, Iss 7, p 484 (2021)
Subject
hyperammonemia
inborn errors of metabolism
hemodialysis
infant
Medicine
Language
English
ISSN
2072-6651
Abstract
Acute neonatal hyperammonemia is associated with poor neurological outcomes and high mortality. We developed, based on kinetic modeling, a user-friendly and widely applicable algorithm to tailor the treatment of acute neonatal hyperammonemia. A single compartmental model was calibrated assuming a distribution volume equal to the patient’s total body water (V), as calculated using Wells’ formula, and dialyzer clearance as derived from the measured ammonia time–concentration curves during 11 dialysis sessions in four patients (3.2 ± 0.4 kg). Based on these kinetic simulations, dialysis protocols could be derived for clinical use with different body weights, start concentrations, dialysis machines/dialyzers and dialysis settings (e.g., blood flow QB). By a single measurement of ammonia concentration at the dialyzer inlet and outlet, dialyzer clearance (K) can be calculated as K = QB∙[(Cinlet − Coutlet)/Cinlet]. The time (T) needed to decrease the ammonia concentration from a predialysis start concentration Cstart to a desired target concentration Ctarget is then equal to T = (−V/K)∙LN(Ctarget/Cstart). By implementing these formulae in a simple spreadsheet, medical staff can draw an institution-specific flowchart for patient-tailored treatment of hyperammonemia.