학술논문

Smartphone camera oximetry in an induced hypoxemia study
Document Type
article
Source
npj Digital Medicine. 5(1)
Subject
Health Services and Systems
Health Sciences
Lung
Clinical Research
Detection
screening and diagnosis
4.2 Evaluation of markers and technologies
Respiratory
Good Health and Well Being
Health services and systems
Language
Abstract
Hypoxemia, a medical condition that occurs when the blood is not carrying enough oxygen to adequately supply the tissues, is a leading indicator for dangerous complications of respiratory diseases like asthma, COPD, and COVID-19. While purpose-built pulse oximeters can provide accurate blood-oxygen saturation (SpO2) readings that allow for diagnosis of hypoxemia, enabling this capability in unmodified smartphone cameras via a software update could give more people access to important information about their health. Towards this goal, we performed the first clinical development validation on a smartphone camera-based SpO2 sensing system using a varied fraction of inspired oxygen (FiO2) protocol, creating a clinically relevant validation dataset for solely smartphone-based contact PPG methods on a wider range of SpO2 values (70-100%) than prior studies (85-100%). We built a deep learning model using this data to demonstrate an overall MAE = 5.00% SpO2 while identifying positive cases of low SpO2