Titanium Fastener Utilization During HeartMate 3 Left Ventricular Assist Device Implantation
Document Type
Academic Journal
Source
Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery. Sep 01, 2018 13(5):361-364
Subject
Language
English
ISSN
1556-9845
Abstract
OBJECTIVE: The objective of this study was to investigate the feasibility of using titanium fasteners for securement of the inflow sewing ring during HeartMate 3 implantation. The secondary objective was to compare cardiopulmonary bypass and total operative times between the titanium fastener and hand-tied knot groups. METHODS: Clearance between the sewing ring and the HeartMate 3 device was assessed in vitro. Thirty-one patients undergoing HeartMate 3 implantation via median sternotomy at a single center from April 2017 to February 2018 were reviewed. The sewing ring was secured with hand-tied knots (n = 18) or titanium fasteners (n = 13). Cannulation strategy and implantation technique were otherwise identical between groups. Central arterial and venous cannulation was performed for cardiopulmonary bypass. The left ventricular apex was cored, and the sewing ring was attached with hand-tied knots or titanium fasteners. RESULTS: There was adequate clearance for the titanium fastener to secure the inflow sewing ring and then connect to the HeartMate 3 in vitro. The inflow sewing ring was successfully secured during HeartMate 3 implantation in the titanium fastener group. Cardiopulmonary bypass time was 75 and 92 minutes for the titanium fastener and hand-tied groups, respectively (P < 0.03). Total operative time was 177 and 193 minutes for the titanium fastener and hand-tied groups, respectively (P = 0.513). CONCLUSIONS: The inflow sewing ring of the HeartMate 3 was efficiently secured using titanium fasteners. Titanium fasteners resulted in shorter cardiopulmonary bypass times compared with the hand-tied group. There was no difference in total operative time.