Precision and accuracy of FEV.sub.1 measurements from the Vitalograph copd-6 mini-spirometer in a healthy Ugandan population
Document Type
Report
Source
PLoS ONE. June 28, 2021, Vol. 16 Issue 6, e0253319
Subject
Uganda
Language
English
ISSN
1932-6203
Abstract
Objective Evaluate the accuracy and precision of the copd-6 mini-spirometer for FEV.sub.1 in a rural Ugandan population. Methods In a cross-sectional study, 171 smallholder farmers performed spirometry with copd-6, and a diagnostic-quality spirometer. Results and discussion The copd-6 underestimated FEV.sub.1 at low flows and overestimated FEV.sub.1 at high flows. Across all participants, the device slightly overestimated FEV.sub.1 by 0.04 [0.02; 0.06] L. Calibration data showed similar patterns. Conclusion The copd-6 could be considered as an affordable tool for research on lung function impairment in resource-constrained settings. However, further validation in a study population with obstructive lung disease is needed.
Author(s): Wajd Abbas Hassan Hansen 1,2, Vivi Schlünssen 2,3, Erik Jørs 4,5, Daniel Sekabojja 6, John C. Ssempebwa 7, Ruth Mubeezi 7, Philipp Staudacher 8,9, Samuel Fuhrimann 10, Martin Rune [...]