Personal Dosimetry in Direct Pulsed Photon Fields With the Dosepix Detector
Document Type
Periodical
Source
IEEE Transactions on Nuclear Science IEEE Trans. Nucl. Sci. Nuclear Science, IEEE Transactions on. 69(12):2330-2334 Dec, 2022
Subject
Nuclear Engineering
Bioengineering
Photonics
Detectors
Dosimetry
Uncertainty
Phantoms
Pixel
Electron tubes
Active personal dosimetry
Dosepix
hybrid pixel detector
pulsed photon fields
Language
ISSN
0018-9499
1558-1578
Abstract
First investigations regarding dosimetric properties of the hybrid, pixelated, photon-counting Dosepix detector in the direct beam of a pulsed photon field (RQR8) for the personal dose equivalent $H\mathrm {_{p}(10)}$ are presented. The influence quantities such as pulse duration and dose rate were varied, and their responses were compared to the legal limits provided in PTB-A 23.2. The variation of pulse duration at a nearly constant dose rate of about 3.7 Sv/h shows a flat response around 1.0 from 3.6 s down to 2 ms. A response close to 1.0 is achieved for dose rates from 0.07 to 35 Sv/h for both pixel sizes. Above this dose rate, the large pixels ( $220~\mu \text{m}$ edge length) are below the lower limit. The small pixels ( $55~\mu \text{m}$ edge length) stay within limits up to 704 Sv/h. The count rate linearity is compared to previous results, confirming the saturating count rate for high dose rates.