학술논문

Simultaneous measurement of neutron and gamma-ray radiation levels from a TRIGA reactor core using silicon carbide semiconductor detectors
Document Type
Conference
Source
1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255) Nuclear science and medical imaging Nuclear Science Symposium, 1998. Conference Record. 1998 IEEE. 3:2136-2141 vol.3 1998
Subject
Nuclear Engineering
Power, Energy and Industry Applications
Fields, Waves and Electromagnetics
Engineered Materials, Dielectrics and Plasmas
Neutrons
Gamma ray detection
Gamma ray detectors
Silicon carbide
Gamma rays
Radiation detectors
Inductors
Length measurement
Fuels
Monitoring
Language
ISSN
1082-3654
Abstract
The ability of a SiC detector to measure neutron and gamma radiation levels in a TRIGA reactor's mixed neutron/gamma field was demonstrated. Linear responses to an epicadmium neutron fluence rate (up to 3/spl times/10/sup 7/ cm/sup -2/ s/sup -1/) and to a gamma dose rate (0.6-234 krad-Si h/sup -1/) were obtained with the detector. Axial profiles of the reactor core's neutron and gamma-ray radiation levels were successfully generated through sequential measurements along the length of the core. The SiC detector shows a high level of precision for both neutrons and gamma rays in high-intensity radiation environments -1.9% for neutrons and better than 0.6% for gamma rays. These results indicate that SiC detectors are well suited for applications such as spent fuel monitoring where measurements in mixed neutron/gamma fields are desired.