GEANT4 simulations and experimental measurements of absolute source activity using modified sum-peak method
Document Type
Conference
Source
2015 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2015 IEEE. :1-3 Oct, 2015
Subject
Bioengineering
Components, Circuits, Devices and Systems
Nuclear Engineering
Signal Processing and Analysis
Detectors
Energy measurement
Isotopes
Government
Calibration
Gamma-rays
Radioactive decay
Language
Abstract
In the present work, we have made Monte Carlo simulations of the response of 1″ × 1″ LaCb(Ce) and 1″ × 1″ LaBn(Ce) scintillation detectors using the radioactive sources emitting two gamma rays in cascade, namely, 60 Co, 94 Nb, 46 Sc and 24 Na for different values of source-detector separation. These simulations were done using GEANT4 toolkit that include radioactive decay module and general particle source (GPS) module. Energy spectra were generated for each of these double gamma emitters and absolute activities were calculated using conventional sum-peak method and modified sum-peak method proposed very recently. In order to validate the simulated results, we have carried out experimental investigations using a calibrated 60 Co source. The measured activities are found to be in very good agreement with the simulated activities.