Comparison of glucose monitoring between Freestyle Libre Pro and iPro2 in patients with diabetes mellitus
Document Type
article
Source
Journal of Diabetes Investigation, Vol 10, Iss 3, Pp 851-856 (2019)
Subject
Nocturnal hypoglycemia
Postprandial hyperglycemia
Self‐management
Diseases of the endocrine glands. Clinical endocrinology
RC648-665
Language
English
ISSN
2040-1124
2040-1116
Abstract
Abstract Aims/Introduction Flash and continuous glucose monitoring systems are becoming prevalent in clinical practice. We directly compared a flash glucose monitoring system (FreeStyle Libre Pro [FSL‐Pro]) with a continuous glucose monitoring system (iPro2) in patients with diabetes mellitus. Materials and Methods Glucose concentrations were simultaneously measured using the FSL‐Pro, iPro2 and self‐monitoring blood glucose in 10 patients with diabetes mellitus, and agreement among them was assessed. Results Parkes error grid analysis showed that the 92.9 and 7.1% of glucose values measured using the FSL‐Pro fell into areas A and B, respectively, and that 96.3, 2.8 and 0.9% of those determined using iPro2 fell into areas A, B and C, respectively. The median absolute relative differences compared with self‐monitoring blood glucose were 8.1% (3.9–12.7%) and 5.0% (2.6–9.1%) for the FSL‐Pro and iPro2, respectively. Analysis of 5,555 paired values showed a close correlation between FSL‐Pro and iPro2 glucose values (ρ = 0.96, P