Confirming X-ray Parametric Down Conversion by Time-Energy Correlation
Document Type
Working Paper
Source
Subject
Quantum Physics
Physics - Optics
Language
Abstract
We present measurements of X-ray Parametric Down Conversion at the Advanced Photon Source synchrotron facility. Using an incoming pump beam at 22 keV, we observe the simultaneous, elastic emission of down-converted photon pairs generated in a diamond crystal. The pairs are detected using high count rate silicon drift detectors with low noise. Production by down-conversion is confirmed by measuring time-energy correlations in the detector signal, where photon pairs within an energy window ranging from 10 to 12 keV are only observed at short time differences. By systematically varying the crystal misalignment and detector positions, we obtain results that are consistent with the constant total of the down-converted signal. Our maximum rate of observed pairs was 130 /hour, corresponding to a conversion efficiency for the down-conversion process of $5.3 \pm 0.5 \times 10^{-13}$.