Development of wide range photon detection system for muonic X-ray spectroscopy
Document Type
Working Paper
Source
Subject
Physics - Instrumentation and Detectors
Nuclear Experiment
Language
Abstract
We have developed a photon detection system for muonic X-ray spectroscopy. The detector system consists of high-purity germanium detectors with BGO Compton suppressors. The signals from the detectors are readout with a digital acquisition system. The absolute energy accuracy, energy and timing resolutions, photo-peak efficiency, the performance of the Compton suppressor, and high count rate durability are studied with standard $\gamma$-ray sources and in-beam experiment using $^{27}\mathrm{Al}(p, \gamma){}^{28}\mathrm{Si}$ resonance reaction. The detection system was demonstrated at Paul Scherrer Institute. A calibration method for a photon detector at a muon facility using muonic X-rays of $^{197}$Au and $^{209}$Bi is proposed.