New Experimentally Observable Gamma-ray Emissions from 241Am Nuclear Decay
Document Type
Working Paper
Author
Koehler, Katrina E.; Yoho, Michael D.; Carpenter, Matthew H.; Croce, Mark P.; Mercer, David J.; Smith, Chandler M.; Tollefson, Aidan D.; Vo, Duc T.; Famiano, Michael A.; Becker, Daniel T.; Gard, Johnathon D.; Mates, J. A. B.; Ortiz, Nathan J.; Ullom, Joel N.; Wessels, Abigail L.; Bennett, Douglas A.; Schmidt, Daniel R.; Swetz, Daniel S.; Vale, Leila R.
Source
Subject
Language
Abstract
With the high resolution of microcalorimeter detectors, previously unresolvable gamma-ray lines are now clearly resolvable. A careful measurement of 241Am decay with a large microcalorimeter array has yielded never before seen or predicted gamma lines at 207.72 +/- 0.02 keV and 208.21 +/- 0.01 keV. These have been made possible because of new microwave-multiplexing readout and improved analysis algorithms for microcalorimeters.
Comment: 5 pages, 3 figures, 2 tables