학술논문

TIARA: a large solid angle silicon array for direct reaction studies with radioactive beams
Document Type
Working Paper
Source
Nucl.Instrum.Meth.A614:439-448,2010
Subject
Nuclear Experiment
Language
Abstract
A compact, quasi-4pi position sensitive silicon array, TIARA, designed to study direct reactions induced by radioactive beams in inverse kinematics is described here. The Transfer and Inelastic All-angle Reaction Array (TIARA) consists of 8 resistive charge division detectors forming an octagonal barrel around the target and a set of double-sided silicon-strip annular detectors positioned at each end of the barrel. The detector was coupled to the -ray array EXOGAM and the spectrometer VAMOS at the GANIL Laboratory to demonstrate the potential of such an apparatus with radioactive beams. The 14N(d,p)15N reaction, well known in direct kinematics, has been carried out in inverse kinematics for that purpose. The observation of the 15N ground state and excited states at 7.16 and 7.86 MeV is presented here as well as the comparison of the measured proton angular distributions with DWBA calculations. Transferred l-values are in very good agreement with both theoretical calculations and previous experimental results obtained in direct kinematics.
Comment: 12 figures, submitted to NIM A