학술논문

Development and validation of a UPLC-MS/MS method to quantitate anti-PD1 monoclonal antibody (Toripalimab), and comparison with electrochemiluminescence immunoassay
Document Type
Article
Source
In Journal of Pharmaceutical and Biomedical Analysis 5 February 2022 209
Subject
Language
ISSN
0731-7085
Abstract
Toripalimab, a humanized IgG4 monoclonal antibody (mAb) against programmed death receptor-1, is being extensively studied to treat various malignancies. At present, there is no complete methodology reported for quantifying toripalimab, except for an electrochemiluminescence immunoassay (ECLIA) mentioned in several clinical studies. Therefore, a sensitive and robust ultra-high performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) was developed to accurately detect toripalimab levels, compared with the ECLIA. Plasma samples were pretreated by a five-step process, encompassing denaturation, reduction, alkylation, enzymatic hydrolysis and quenching. And a unique, sensitive and stable enzymatic peptide (ASGYTFTDYEMHWVR) selected as surrogate of toripalimab was eluted and monitored by UPLC-MS/MS system with the linear range of 5.0375–201.5 μg/mL. After fully validated, the UPLC-MS/MS method was applied to determine 77 plasma samples from 29 patients in a phase I clinical trial, and compared with ECLIA based on 56 samples. Wilcoxon paired samples test showed toripalimab levels by UPLC-MS/MS were significantly higher than that by ECLIA (p