Murine Heart Volume: Numerical Comparison and Calibration of Conductance Catheter Models.
Document Type
Article
Source
IEEE Transactions on Biomedical Engineering. Sep2014, Vol. 61 Issue 9, p2396-2405. 10p.
Subject
*CATHETERIZATION
*CARDIAC catheterization
*FINITE element method
*ELECTROMAGNETISM
*MEDICAL simulation
*LABORATORY mice
*CALIBRATION
Language
ISSN
0018-9294
Abstract
A full set of finite-element method (FEM) studies of the catheter within a cylindrical cuvette and within an elliptical cuvette are presented along with novel insight on the fundamental electromagnetic properties of the catheter. An in vitro experiment with modified small mouse pressure–volume catheters was conducted and the results are presented as a validation of the FEM models. In addition, sensitivity analysis on the electrode size and position is conducted and the results allow for a novel calibration factor based on catheter geometry to be presented. This calibration factor is used in conjunction with Wei’s conductance volume equations to reduce the average measured error in cuvette volume measurements from 26.5% to 5%. [ABSTRACT FROM PUBLISHER]