Evaluation of Reconstruction Algorithms to Validate the NEMA Phantom Results in Clinical Scenario - A Comparative Study Using Time-of-Flight versus Non-Time-of-Flight Positron Emission Tomography Imaging.
Document Type
Article
Source
Indian Journal of Nuclear Medicine. Apr-Jun2022, Vol. 37 Issue 2, p113-120. 8p.
Subject
*POSITRON emission tomography
*IMAGE analysis
*COMPUTED tomography
*SIGNAL-to-noise ratio
Language
ISSN
0972-3919
Abstract
Objectives: The objective is to standardize the reconstruction parameters for the time-of-flight (TOF) versus non-TOF positron-emission tomography/computed tomography (PET/CT) imaging data and validation of the same in a clinical setting. Methods: The four spheres (10.0/13.0/17.0/22.0 mm) of the PET phantom (NEMA IQ Nu 2-2001) were filled with four times higher activity of [18F]-NaF than the background (5.3kBq/mL). Imaging (image matrix - 128 x 128 x 47, 2 min, 3D model) was done using two different (TOF/non-TOF) PET scanners. Phantom data were reconstructed in TOF and non-TOF modes for lutetium-yttrium oxyorthosilicate and non-TOF mode for bismuth germanate-based PET scanners. The reconstructed data (by varying iteration and subsets) that provided the best image contrast and signal-to-noise ratio (SNR) were evaluated. The whole-body [18F]-fludeoxyglucose (FDG) PET/CT scans (7-8 frames; 2.0 min/frame) in 16 lymphoma patients were acquired at 60 min after injecting the radioactivity (370.0-444.0 MBq of [18F]-FDG. The clinical PET/CT data were reconstructed using phantom-derived reconstruction parameters and evaluated for image contrast and SNR of the detected lesions. Results: TOF reconstruction at second iteration provided significantly (P ≤ 0.02) higher SNR (20.7) and contrast (contrast recovery coefficient/background variability = 3.21) for the smallest hot lesions (10.0 mm) in the phantom than the non-TOF system. Similarly, in patient data analysis for the selected FDG avid lesions, the SNR values were significantly (P = 0.02) higher (13.3 ± 6.49) in TOF than (11 ± 6.48) in non-TOF system. Further, the small (≤10.0 mm) lesions were seen more distinctly in TOF system. Conclusion: It is thus observed that TOF reconstruction converged faster than the non-TOF, and the applicability of the same may impact the image quality and interpretation in the clinical PET data. The validation of the phantom-based experimental reconstruction parameters to clinical PET imaging data is highly warranted. [ABSTRACT FROM AUTHOR]