Dose distribution verification in intraoperative radiation therapy using an N-isopropyl acrylamide-based polymer gel dosimeter.
Document Type
Article
Source
Journal of Radioanalytical & Nuclear Chemistry. May2020, Vol. 324 Issue 2, p481-488. 8p. 1 Color Photograph, 1 Diagram, 3 Charts, 4 Graphs.
Subject
*POLYMER colloids
*INTRAOPERATIVE radiotherapy
*IONIZATION chambers
*DOSIMETERS
*ACRYLAMIDE
*RADIATION dosimetry
Language
ISSN
0236-5731
Abstract
The purpose of the current study was to verify the dose distribution of an Intrabeam-50 kV IORT system using polymer gel dosimetry technique. Results of dose distribution evaluation using NIPAM polymer gel dosimetry were compared with those measured using an ionization chamber and simulated using MCNPX code. Results showed the calculated gamma index was less than 1 with 2% dose-difference/2 mm distance-to-agreement for comparison between NIPAM and ionization chamber as well as between NIPAM and MCNPX simulation. It was concluded that the NIPAM polymer gel dosimetry is useful for verifying the dose distribution of low energy X-ray IORT technique. [ABSTRACT FROM AUTHOR]