학술논문

Edge-subtraction X-ray ptychographic imaging with pink beam synchrotron radiation and a single photon-counting detector.
Document Type
Article
Source
Scientific Reports. 4/16/2020, Vol. 10 Issue 1, p1-7. 7p.
Subject
*SYNCHROTRON radiation
*PHOTON counting
*GOLD nanoparticles
*MONOCHROMATORS
*ABSORPTION
Language
ISSN
2045-2322
Abstract
We present here a new method of performing X-ray edge-subtraction ptychographic imaging by combining multiple harmonics from an undulator synchtrotron source and an energy discriminating photon counting detector. Conventionally, monochromatic far-field X-ray ptychography is used to perform edge subtraction through the use of multiple monochromatic energy scans to obtain spectral information for a variety of applications. Here, we use directly the undulator spectrum from a synchrotron source, selecting two separate harmonics post sample using the Pixirad-1/Pixie-III detector. The result is two monochromatic images, above and below an absorption edge of interest. The proposed method is applied to obtain Au L-edge subtraction imaging of a Au-Ni grid test sample. The Au L-edge subtraction is particularly relevant for the identification of gold nanoparticles for biomedical applications. Switching the energy scan mechanism from a mechanical monochromator to an electronic detector threshold allows for faster spectral data collection with improved stability. [ABSTRACT FROM AUTHOR]