Pre-treatment 2D and 3D dosimetric verification of volumetric arc therapy. A correlation study between gamma index passing rate and clinical dose volume histogram.
Document Type
Article
Source
PLoS ONE. 8/13/2019, Vol. 14 Issue 8, p1-17. 17p.
Subject
*MONTE Carlo method
*VOLUMETRIC-modulated arc therapy
*RECEIVER operating characteristic curves
*GLEASON grading system
*ENDOMETRIAL cancer
*PROSTATE cancer
*HISTOGRAMS
*PEARSON correlation (Statistics)
Language
ISSN
1932-6203
Abstract
Objectives: To evaluate methods for the pre-treatment verification of volumetric modulated arc therapy (VMAT) based on the percentage gamma passing rate (%GP) and its correlation and sensitivity with percentage dosimetric errors (%DE). Methods: A total of 25 patients with prostate cancer and 15 with endometrial cancer were analysed. The %GP values of 2D and 3D verifications with different acceptance criteria (1%/1 mm, 2%/2 mm, and 3%/3 mm) were obtained using OmniPro and Compass. The %DE was calculated using a planned dose volume histogram (DVH) created in Monaco’s treatment planning system (TPS), which relates radiation dose to tissue and the patient’s predicted dose volume histogram in Compass. Statistical correlation between %GP and %DE was verified using Pearson’s correlation coefficient. Sensitivity was calculated based on the receiver operating characteristics (ROC) curve. Plans were calculated using Collapsed Cone Convolution and the Monte Carlo algorithm. Results: The t-test results of the planned and estimated DVH showed that the mean values were comparable (P > 0.05). For the 3%/3 mm criterion, the average %GP was acceptable for the prostate and endometrial cancer groups, with average rates of 99.68 ± 0.49% and 99.03 ± 0.59% for 2D and 99.86 ± 0.39% and 99.53 ± 0.44% for 3D, respectively. The number of correlations was poor for all analysed data. The mean Pearson’s R-values for prostate and endometrial cancer were < 0.45 and < 0.43, respectively. The area under the ROC curve for the prostate and endometrial cancer groups, was lower than 0.667. Conclusions: Analysis of the %GP versus %DE values revealed only weak correlations between 2D and 3D verifications. DVH results obtained using the Compass system will be helpful in confirming that the analysed plans respect dosimetric constraints. [ABSTRACT FROM AUTHOR]