학술논문

How does motion affect GABA-measurements? Order statistic filtering compared to conventional analysis of MEGA-PRESS MRS.
Document Type
Article
Source
PLoS ONE. 5/16/2017, Vol. 12 Issue 5, p1-20. 20p.
Subject
*AMINOBUTYRIC acid
*L-moments
*GABAERGIC neurons
*KINETIC energy
*MOTION detectors
Language
ISSN
1932-6203
Abstract
Purpose: The aim of this study was to evaluate two post-processing techniques applied to MRS MEGA-PRESS data influenced by motion-induced artifacts. In contrast to the conventional averaging technique, order statistic filtering (OSF) is a known method for artifact reduction. Therefore, this method may be suitable to incorporate in the GABA quantification. Methods: Twelve healthy volunteers were scanned three times using a 3 T MR system. One measurement protocol consisted of two MEGA-PRESS measurements, one reference measurement and one measurement including head motions. The resulting datasets were analyzed with the standard averaging technique and with the OSF-technique in two schemes; filtering phase cycles ‘RAW PC’ and filtering dynamics ‘RAW Dyn’. Results: The datasets containing artifacts resulted in an underestimation of the concentrations. There was a trend for the OSF-technique to compensate for this reduction when quantifying SNR-intense signals. However, there was no indication that OSF improved the estimated GABA concentrations. Moreover, when only considering the reference measurements, the OSF technique was equally as effective as averaging, which suggests that the techniques are interchangeable. Conclusion: OSF performed equally well as the conventional averaging technique for low-SNR signals. For high-SNR signals, OSF performed better and thus could be considered for routine usage. [ABSTRACT FROM AUTHOR]