학술논문

Selective cloud point extraction and preconcentration of trace amounts of cadmium prior to flame atomic absorption spectrometric determination
Document Type
article
Source
Journal of the Brazilian Chemical Society. December 2010 21(12)
Subject
cloud point preconcentration
cadmium ion determination
FI-FAAS
Trion X-114
water samples
ore
Language
English
ISSN
0103-5053
Abstract
Cloud point extraction (CPE) is proposed as a separation and preconcentration procedure for determination of cadmium in water samples by flow injection flame atomic absorption spectrometry (FI-FAAS). The cadmium ion in aqueous solution was converted to [CdI4]2- and its charge was neutralized through ion-pair formation with methyltrioctylammonium cation in the presence of Triton X-114 as a non ionic surfactant. Phase separation was achieved by heating the mixture to 80 ºC for 10 min. The surfactant-rich phase was diluted with 300 µL of ethanol and 100 µL of it was analysed by FI-FAAS. The parameters affecting the ion-pair formation and extraction steps were optimized. Under optimum condition, preconcentration of 35 mL of sample in the presence of 0.04% (v/v) of Triton X- 114 led to a detection limit of 0.3 µg L-1 of Cd(II). Analytical calibration curve was linear up to 250 µg L-1 of cadmium and relative standard deviation was lower than 4% at 100 µg L-1 level. The procedure affords recoveries in the range of 95-102%. The procedure was successfully applied to the determination of Cd in tap water, sea water and certified reference copper concentrate.