학술논문

Safety and efficacy results of the Flow Redirection Endoluminal Device (FRED) stent system in the treatment of intracranial aneurysms: US pivotal trial.
Document Type
article
Source
Journal of neurointerventional surgery. 14(6)
Subject
Humans
Intracranial Aneurysm
Treatment Outcome
Embolization
Therapeutic
Prospective Studies
Stents
Stroke
Endovascular Procedures
aneurysm
flow diverter
intervention
technology
Comparative Effectiveness Research
Neurosciences
Clinical Research
Clinical Trials and Supportive Activities
Bioengineering
Cardiovascular
Language
Abstract
ObjectiveTo evaluate the safety and effectiveness of the Flow Redirection Endoluminal Device (FRED) flow diverter in support of an application for Food and Drug Administration approval in the USA.Methods145 patients were enrolled in a prospective, single-arm multicenter trial. Patients with aneurysms of unfavorable morphology for traditional endovascular therapies (large, wide-necked, fusiform, etc) were included. The trial was designed to demonstrate non-inferiority in both safety and effectiveness, comparing trial results with performance goals (PGs) established from peer-reviewed published literature. The primary safety endpoint was death or major stroke (National Institutes of Health Stroke Scale score ≥4 points) within 30 days of the procedure, or any major ipsilateral stroke or neurological death within the first year. The primary effectiveness endpoint was complete occlusion of the target aneurysm with ≤50% stenosis of the parent artery at 12 months after treatment, and in which an alternative treatment of the target intracranial aneurysm had not been performed.Results145 patients underwent attempted placement of a FRED device, and one or more devices were placed in all 145 patients. 135/145 (93%) had a single device placed. Core laboratory adjudication deemed 106 (73.1%) of the aneurysms large or giant. A safety endpoint was experienced by 9/145 (6.2%) patients, successfully achieving the safety PG of 46% aneurysm occlusion was also achieved, with the effectiveness endpoint being met in 80/139 (57.6%) CONCLUSION: As compared with historically derived performance benchmarks, the FRED flow diverter is both safe and effective for the treatment of appropriately selected intracranial aneurysms.Clinical registration numberNCT01801007.