Clinical Evaluation of BD Veritor SARS-CoV-2 Point-of-Care Test Performance Compared to PCR-Based Testing and versus the Sofia 2 SARS Antigen Point-of-Care Test
Document Type
article
Source
Journal of Clinical Microbiology. 59(1)
Subject
Biomedical and Clinical Sciences
Clinical Sciences
Clinical Research
Biodefense
Emerging Infectious Diseases
Infectious Diseases
Lung
Prevention
Vaccine Related
Good Health and Well Being
Adult
Antigens
Viral
COVID-19
COVID-19 Testing
Coronavirus Nucleocapsid Proteins
Female
Humans
Immunoassay
Male
Middle Aged
Nasopharynx
Oropharynx
Point-of-Care Testing
Polymerase Chain Reaction
SARS-CoV-2
Sensitivity and Specificity
Spike Glycoprotein
Coronavirus
Veritor test
point-of-care test
Sofia 2 test
Biological Sciences
Agricultural and Veterinary Sciences
Medical and Health Sciences
Microbiology
Clinical sciences
Medical microbiology
Language
Abstract
The clinical performance of the BD Veritor System for Rapid Detection of SARS-CoV-2 nucleocapsid antigen (Veritor), a chromatographic immunoassay used for SARS-CoV-2 point-of-care testing, was evaluated using nasal specimens from individuals with COVID-19 symptoms. Two studies were completed to determine clinical performance. In the first study, nasal specimens and either nasopharyngeal or oropharyngeal specimens from 251 participants with COVID-19 symptoms (≤7 days from symptom onset [DSO], ≥18 years of age) were utilized to compare Veritor with the Lyra SARS-CoV-2 PCR assay (Lyra). In the second study, nasal specimens from 361 participants with COVID-19 symptoms (≤5 DSO, ≥18 years of age) were utilized to compare performance of Veritor to that of the Sofia 2 SARS Antigen FIA test (Sofia 2). The positive, negative, and overall percent agreement (PPA, NPA, and OPA, respectively) were the primary outcomes. In study 1, the PPA for Veritor, compared to Lyra, ranged from 81.8 to 87.5% across the 0 to 1 and 0 to 6 DSO ranges. In study 2, Veritor had PPA, NPA, and OPA values of 97.4, 98.1, and 98.1%, respectively, with Sofia 2. Discordant analysis showed one Lyra positive missed by Veritor and five Lyra positives missed by Sofia 2; one Veritor positive result was negative by Lyra. Veritor met FDA emergency use authorization (EUA) acceptance criteria for SARS-CoV-2 antigen testing for the 0 to 5 and 0 to 6 DSO ranges (PPA values of 83.9% and 82.4%, respectively). Veritor and Sofia 2 showed a high degree of agreement for SARS-CoV-2 detection. The Veritor test allows for more rapid COVID-19 testing utilizing easy-to-collect nasal swabs but demonstrated