[단신] 즉발감마선 계측시스템의 반사체를 이용한 열중성자 효율증대 연구
Document Type
Academic Journal
Source
분석과학. 2003-10 16(5):426-429
Subject
Language
Korean
ISSN
1225-0163
2288-8985
Abstract
Neutron induced prompt gamma-ray spectroscopy (NIPS) system equipped with a (252)^Cf neutron source and a n-type coaxial HPGe detector was installed for the quantitative analysis of aqueous samples in KAERI, Korea. Since the thermal neutron flux for the (252)^Cf neutron source is relatively low compared to that for the reactor, the use of a thermal neutron reflector in the NIPS system may lead to improved results. The enhancement by using various reflectors was carried out by comparing the Cl peak with or without a cadmium plate between sample and the (252)^Cf source. The use of pyrolitic graphite as a reflector provided a good result.