Appropriate parameters of the ordered-subset expectation maximization algorithm on measurement of myocardial blood flow and oxygen consumption with 11C-acetate PET
Document Type
Academic Journal
Source
Nuclear Medicine Communications. Feb 01, 2012 33(2):130-138
Subject
Language
English
ISSN
0143-3636
Abstract
OBJECTIVE: The purpose of this study was to evaluate the appropriate parameters of a filter and of subsets (S) and iterations (I) of the ordered-subset expectation maximization (OSEM) algorithm in C-acetate PET. METHODS: A Hanning filter (HF) and a Gaussian filter (GF) were selected for filtered back-projection (FBP) and the OSEM algorithm, respectively. After evaluation of the optimal HF size, the GF size was optimized using healthy volunteers (HV). Myocardial blood flow (MBF) and oxygen consumption (kmono) values were calculated by combining 4S, 16S, or 28S with 2I, 4I, 6I, or 8I of the OSEM (MBFOSEM and kmonoOSEM, respectively) in eight HV and eight coronary artery disease (CAD) patients. These MBFOSEM and kmonoOSEM values were compared with those obtained using FBP (MBFFBP and kmonoFBP, respectively). RESULTS: Optimal HF and GF (10.0GF) sizes for the FBP and OSEM algorithms, respectively, were 10.0 mm full-width resolution at half-maximum. MBFOSEM was changed by modifying the parameters of the OSEM algorithm. The best correlations were between MBFFBP and MBFOSEM, with 28S6I and 10.0GF for HV patients and 28S8I for CAD patients. However, the MBFOSEM with 28S8I was significantly different from MBFFBP at the global myocardium in HV. The kmonoOSEM with 28S6I was not significantly different from kmonoFBP in HV or CAD patients. CONCLUSION: Appropriate parameters are 28S6I with a 10.0GF on the MBFOSEM and kmonoOSEM measurement using C-acetate. Diagnostic performance will improve using noiseless, artifact-reduction images, and accurate quantitative values that are provided by the OSEM algorithm with the appropriate parameters.