학술논문

Detector response in time-of-flight mass spectrometry at high pulse repetition frequencies
Document Type
Report
Source
Review of Scientific Instruments. 64(8)
Subject
Instrumentation And Photography
Language
English
ISSN
0034-6748
Abstract
Dead time effects in chevron configured dual microchannel plates (MCPs) are investigated. Response times are determined experimentally for one chevron-configured dual MCP-type detector and two discrete dynode-type electron multipliers with 16 and 23 resistively divided stages. All of these detectors are found to be suitable for time-of-flight mass spectrometry (TOF MS), yielding 3-6-ns (FWHM) response times triggered on a single ion pulse. It is concluded that, unless there are viable solutions to overcome dead time disadvantages for continuous dynode detectors, suitable discrete dynode detectors for TOF MS appear to have a significant advantage for high repetition rate operation.