학술논문

Portable Laser Doppler Flowmeter for Microcirculation Detection
Document Type
Article
Source
Biomedical Engineering Letters (BMEL), 3(2), pp.109-114 Jun, 2013
Subject
의공학
Language
English
ISSN
2093-985X
2093-9868
Abstract
Purpose The microcirculatory system of the human body is closely related to its physiological conditions. The laser Doppler flowmeter can detect symptoms through variations in the microcirculation; however, the present commercial Doppler flowmeter is very expensive, heavy, and inconvenient to carry. Therefore, an inexpensive, small, and portable laser Doppler flowmeter is proposed. Methods In this study, we designed a laser Doppler flowmeter transmitting receiving module based on the fundamental principles of laser Doppler, using a common computer for flow calculation. A self-designed correction system was used to validate the portable laser Doppler flowmeter. Results The coefficient of correlation between the mean frequency offset <ω> of the flowmeter and the actual mimicked blood flow rate reached a high value of R = 0.99, and the variations in microcirculation flow could be accurately measured. Conclusions The results proved that the low-cost portable laser Doppler flowmeter can be used extensively in early symptom detection for disease prevention and control.