Evaluation of energy correction algorithm for signals of PET in heavy-ion cancer therapy device
Document Type
Article
Source
Nuclear Engineering and Technology, 52(1), pp.101-108 Jan, 2020
Subject
원자력공학
Language
English
ISSN
2234-358X
1738-5733
Abstract
In order to solve the contradiction between requirements of high sampling rate for acquiring accurate energy information of pulses and large amount of data to be processed timely, the method with an algorithm to correct errors caused by reducing the sampling rate is normally used in front-end read-out system, which is conductive to extract accurate energy information from digitized waveform of pulse. The functions and effects of algorithms, which mainly include polynomial fitting with different fitting times, double exponential function fitting under different sampling modes, and integral area algorithm, are analyzed and evaluated, and some meaningful results is presented in this paper. The algorithm described in the paper has been used preliminarily in a prototype system of Positron Emission Tomography (PET) for heavy-ion cancer therapy facility.