Comparison of the Apnea-Hypopnea Index Determined by a Peripheral Arterial Tonometry-Based Device With That Determined by Polysomnography ― Results From a Multicenter Study ―
Document Type
Journal Article
Source
Circulation Reports. 2020, 2(11):674
Subject
Cardiovascular disease
Polygraphy
Sleep disordered breathing
Sympathetic nervous activity
Language
English
ISSN
2434-0790
Abstract
Methods and Results:We enrolled 120 patients undergoing overnight polysomnography at 6 Japanese centers. A PAT-based device was used simultaneously with polysomnography. Polysomnography recordings were scored centrally by a technician in a blinded manner. PAT-based device recordings were scored using an automatic algorithm. There was a strong correlation between pAHI and AHI-PSG (r=0.896; P<0.001) with acceptable agreement. The strong correlation between pAHI and AHI-PSG was observed in patients with CVD (n=55; P=0.849; P<0.001) and without CVD (n=65; r=0.927; P<0.001). The presence or absence of CVD did not affect the relationship between pAHI and AHI-PSG (P=0.225).