학술논문

Basic study of urinary oxalic acid measurement using capillary electrophoresis / キャピラリー電気泳動法を用いた尿中シュウ酸の測定法の基礎的検討
Document Type
Journal Article
Source
医学検査 / Japanese Journal of Medical Technology. 2020, 69(2):179
Subject
capillary electrophoresis
ion chromatograph
urinary oxalic acid
イオンクロマト法
キャピラリー電気泳動法
尿中シュウ酸
Language
Japanese
ISSN
0915-8669
2188-5346
Abstract
We performed basic evaluation studies of the urinary oxalic acid measurement by capillary electrophoresis. This method showed CVs of within-run and between-day precisions of less than 6.0%, and the relative error of dilution linearity was good. The recovery test also showed good results. No effects of hydrochloric acid or ascorbic acid on the urinary oxalic acid measurement were detected. In addition, the correlation of urine samples with high-performance ion chromatography and capillary electrophoresis was good (y = 0.971x + 0.380, r = 0.997, n = 50). This study revealed that this method using capillary electrophoresis is applicable in routine testing.