Pacemaker-induced Metallic Artifacts in Coronary Computed Tomography Angiography: Clinical Feasibility of Single Energy Metal Artifact Reduction Technique / ペースメーカリードが冠動脈CT 血管造影に与える影響と金属アーチファクト低減再構成法の有用性
Document Type
Journal Article
Source
日本放射線技術学会雑誌 / Japanese Journal of Radiological Technology. 2017, 73(6):460
Subject
coronary artery
coronary computed tomography angiography
metallic artifact
pacemaker
Language
Japanese
ISSN
0369-4305
1881-4883
Abstract
Background: Coronary computed tomography angiography (CCTA) in patients with pacemaker suffers from metallic lead-induced artifacts, which often interfere with accurate assessment of coronary luminal stenosis. The purpose of this study was to assess a frequency of the lead-induced artifacts and artifact-suppression effect by the single energy metal artifact reduction (SEMAR) technique. Methods: Forty-one patients with a dual-chamber pacemaker were evaluated using a 320 multi-detector row CT (MDCT). Among them, 22 patients with motion-free full data reconstruction images were the final candidates. Images with and without the SMEAR technique were subjectively compared, and the degree of metallic artifacts was compared. Results: On images without SEMAR, severe metallic artifacts were often observed in the right coronary artery (#1, #2, #3) and distal anterior descending branch (#8). These artifacts were effectively suppressed by SEMAR, and the luminal accessibility was significantly improved in #3 and #8. Conclusion: While pacemaker leads often cause metallic-induced artifacts, SEMAR technique reduced the artifacts and significantly improved the accessibility of coronary lumen in #3 and #8.