Characteristics of the Z-resolution Measurement in the Digital Breast Tomosynthesis / ディジタルブレストトモシンセシスのZ軸分解能測定における諸特性
Document Type
Journal Article
Source
日本放射線技術学会雑誌 / Japanese Journal of Radiological Technology. :2023-1395
Subject
Z-resolution
bead method
mammography
tomosynthesis
Language
Japanese
ISSN
0369-4305
1881-4883
Abstract
Purpose: Digital breast tomosynthesis (DBT) offers a choice of image acquisition conditions. The object of observation is the entire detector in the full depth (Z-axis) direction. In this study, we investigated the effect of each acquisition condition on Z-resolution with a micro-metallic sphere (bead) as an imaged object in DBT images. Method: DBT images were acquired in two imaging modes with different angular ranges, varying with the height of the bead, the bead placement position, and the bead diameter. The depth profile of the bead image was obtained, and the Z-resolution was defined by its full width at half maximum. The Z-resolution was compared with each acquisition condition. Result: The Z-resolution values decreased with increasing height from the breast support and decreasing bead diameter. The Z-resolution values were larger at the edges than at the detector center. Conclusion: This study shows that the bead diameter and the geometry of the bead and detector affect the Z-resolution measurement.