Accuracy Evaluation of Air Kerma-area Product of Over-couch-type X-ray Fluoroscopic System / オーバーテーブル式X線透視装置の面積線量表示値の精度評価
Document Type
Journal Article
Source
日本放射線技術学会雑誌 / Japanese Journal of Radiological Technology. 2022, 78(11):1349
Subject
air kerma
air kerma-area product
over-couch fluoroscopy
Language
Japanese
ISSN
0369-4305
1881-4883
Abstract
Purpose: The purpose of this study was to evaluate the accuracy of incident air kerma (Ka,r) and air kerma-area product (PKA) displayed on over-couch-type X-ray fluoroscopic systems by comparing them with the measured values. Methods: An ionizing chamber was placed at the patient entrance reference point to measure the Ka,r. The PKA was calculated by multiplying the Ka,r by the irradiation area. These measured values were compared with the displayed values. Results: The differences between measured and displayed Ka,r and PKA were less than ±35%, which was the criteria of the Japanese Industrial Standards (JIS). However, the accuracy of the displayed values differed depending on the manufacturer and the device. Conclusion: Although no error exceeding the JIS criteria was observed, it is necessary to understand the characteristics of the X-ray fluoroscopic systems related to displayed dose and to manage the systems by performing dose measurements periodically.