Evaluation of X-ray Shielding Calculation for Medical X-ray Generator / 診断用エックス線装置における遮へい計算評価
Document Type
Journal Article
Source
日本放射線安全管理学会誌 / Japanese Journal of Radiation Safety Management. 2003, 2(1):19
Subject
X-ray room
controlled area
protective shielding
Language
Japanese
ISSN
1347-1503
1884-9512
Abstract
Using the numerical values in the provisions of the law, the proportion evaluated in the safe side was examined. At the time, the following were examined, whether there is no superfluous protection or underestimation in comparison with NCRP and the Law. The transmission rate using in the 1/10 value layer were higher than that calculated in the half value layer in both lead and concrete. For the leakage in X-ray examination room, the effective doses were estimated to be less than the limit of the controlled area of 1.3m Sv/3 months. However, the dose using the law are 3.3 times in maximum in comparison with the NCRP. In the case of QI (Quality Index) =0.5, it was found that the dose were overestimated in 3.6 times in maximum.