Propose for the Benchmark Dose (BD) for the Optimization of Protection in Medical Exposure in General Radiography / 一般撮影の医療被ばくの防護を最適化するためのベンチマークドーズ(BD)の提案
Document Type
Journal Article
Source
日本放射線技術学会雑誌 / Japanese Journal of Radiological Technology. 2018, 74(5):443
Subject
benchmark dose
diagnostic reference level
entrance surface dose
general radiography
medical equipment
Language
Japanese
ISSN
0369-4305
1881-4883
Abstract
The goal of this research was to create the most appropriate index dose for the optimization of protection in medical exposure in general radiography in Kanagawa prefecture. We distributed questionnaires to 272 medical institutions in Kanagawa prefecture. The investigation period was from October 2015 to February 2016. Entrance surface dose (ESD) was used as the index dose. Investigated regions in general radiography were the adult chest, adult abdomen, and infant chest (anterior-posterior projections for all regions). The effective response rate was 35%. ESD was significantly lower with a flat panel detector (FPD) than with computed radiography (CR) in all regions (adult chest and abdomen: p<0.001; infant chest: p<0.05) [e.g., mean (±standard deviation) ESD in the adult chest was 0.16±0.06 mGy with FPD and 0.24±0.10 mGy with CR]. In the infant chest with CR, ESD was significantly higher using a grid (0.15±0.07 mGy) compared to not using a grid (0.10±0.05 mGy; p<0.05). Based on these results, we propose the benchmark dose of each medical equipment, such as adult chest: FPD, 0.2 mGy; CR, 0.3 mGy.