학술논문

High-contrast Visualization of Uterine Three-layer Structure by Fast Spinecho T_2-weighted MR Images : Comparison with Different Parameters / 高速SE法T2強調画像における子宮3層構造描出のための最適パラメータの検討
Document Type
Journal Article
Source
日本放射線技術学会雑誌 / Japanese Journal of Radiological Technology. 2003, 59(4):500
Subject
Fast spin-echo
Original Phantom
T_2-weighted imaging
Uterine three-Layer structure
Language
Japanese
ISSN
0369-4305
1881-4883
Abstract
We performed experimental and clinical MR studies to determine the optimal imaging conditions for the uterine three-layer structure on fast spin-echo T_2-weighted imaging. In the experimental study, we used our original phantom. Imaging studies were carried out with different TRs, TEs, and ETLs, and the contrast-to-noise ratio(CNR)was evaluated. TR was the most important factor for contrast in both the phantom and clinical studies. The longer TR was, the better the contrast became. In the clinical study, TRs of 5500-6000 msec provided clear visualization of the uterine three-layer structure. Longer TRs would be useful for the detection of disease, differential diagnosis, and grading in patients with uterine diseases.