학술논문

A case of angioimmunoblastic T-cell lymphoma exhibiting false high fibrin/fibrinogen degradation product and D-dimer levels / FDPとDダイマーが偽高値を示した血管免疫芽球性T細胞リンパ腫の1症例
Document Type
Journal Article
Source
医学検査 / Japanese Journal of Medical Technology. 2018, 67(1):119
Subject
D-dimer
Dダイマー
angioimmunoblastic T-cell lymphoma
false high value
fibrin/fibrinogen degradation product
hypergammaglobulinemia
フィブリン・フィブリノゲン分解産物
偽高値
血管免疫芽球性T細胞リンパ腫
高γグロブリン血症
Language
Japanese
ISSN
0915-8669
2188-5346
Abstract
In cases of hypergammaglobulinemia due to hematological malignancy, it has been reported that nonspecific reactions are often observed in test items that are affected by the antigen–antibody reactions. We here report on a case of angioimmunoblastic T-cell lymphoma (AITL) that exhibited false high levels of the fibrin/fibrinogen degradation product (FDP) and D-dimer (D-D). The patient was a woman in her 60s. She was diagnosed as having AITL on the basis of multiple lymphadenopathy, rash, and symptoms of liver and lung infiltration. On day 9 of hospitalization, the D-D level was found to be elevated, with the error message “prozone” displayed. As it appeared that the level exceeded the detection range of the device, dilution measurement was performed. As this resulted in a markedly lower level than the theoretical value, a nonspecific reaction was suspected and detailed tests were conducted. As immunoglobulin levels had increased when false high levels of FDP and D-D were detected, and dithiothreitol (DTT) processing decreased measurement values, we estimated that IgM was involved in causing the false high values. Western blotting results also suggested that the diluted measurement values were close to the true levels. As was shown for our patient, when false high levels of FDP and D-D are exhibited with no clinical hemorrhagic or thrombotic symptoms, dilution measurement could be a simple and effective method for determining whether a nonspecific reaction has occurred.