CLINICAL EVALUATION OF PAN-17DX MEMBRANE / PAN17DX膜の臨床評価
Document Type
Journal Article
Source
人工臓器 / Jinko Zoki. 1991, 20(1):12
Subject
antithrombogenicity
biocompatibillity
hemodialysis
polyacrylonitrine membrane
Language
Japanese
ISSN
0300-0818
1883-6097
Abstract
The three polyacrylonitrile membrane dialyzers, PNF-17CX2(CX), PAN-17DX(DX) and Filtral 16(F) were tested for biocompatibility and clinical performance. Eight patients on maintenance hemodialysis took part in this study. They changes of leucocyte counts, platelet counts and C3a at 15 min during the hemodialysis were the same degrees in the three dialyzers. The changes of plateletfactor-4 and activated coagulation time at 15 min during the hemodialysis and residual blood volume within the module fibers after each dialysis session were greater in CX than in other two dialyzers. In conclusion, DX shows the antithrombogenicity better than CX, and as same as F.