학술논문

Misreading of Pressure Measurement by Radiation, and Its Measures / 放射線等による圧力測定異常とその対策
Document Type
Journal Article
Source
真空 / Shinku. 2006, 49(5):302
Subject
Language
Japanese
ISSN
0559-8516
1880-9413
Abstract
In the SPring-8 Storage Ring, the pressure measured by Bayard-Alpert gauge near photon-absorber were not observed aging effects by a beam self-cleaning. As the result of investigations of this phenomenon, three causes of misreading of pressure measurement were found; (1) irradiation of the scattered X-ray to gauge head, (2) a large amount of electronic inflow to gauge head, (3) irradiation of the scattered X-ray to gauge cable. We took measures to the misreading, for example shielding of gauge heads and cables etc, and came to be able to measure pressure within tens of percent errors.