학술논문

A preclinical animal model for evaluating the sealing capacity of covered stent grafts in acute vessel perforation
Document Type
Academic Journal
Source
European Journal of Medical Research. July 29, 2020, Vol. 25 Issue 1
Subject
Stents -- Analysis -- Models -- Health aspects
Polyurethanes -- Analysis -- Models -- Health aspects
Thrombosis -- Models -- Health aspects -- Analysis
Cardiology -- Analysis -- Health aspects -- Models
Polytetrafluoroethylene -- Models -- Analysis -- Health aspects
Animal experimentation -- Analysis -- Health aspects -- Models
Rabbits -- Analysis -- Health aspects -- Models
Language
English
ISSN
0949-2321
Abstract
Background Percutaneous coronary intervention is among the most common therapeutic interventions in cardiology. This procedure may, however, be associated with a rare, though life-threatening complication: acute coronary perforation (CP). CP is primarily treated using covered stents, which are made of bare metal stents with a polytetrafluoroethylene (PTFE) or polyurethane coating. These stents' major limitations include higher rates of thrombus formation and restenosis. Hence, there is a still unmet need for new stents regarding their design and composition. Or, to test new covered stent designs, the rabbit iliac artery has become the best-established animal model. This study sought to present a preclinical animal approach designed to test covered stents that are utilized following vessel perforation. Methods The animal experiments were performed using New Zealand white rabbits, each weighting 3.5-4.5 kg. The animal models described herein relied on the three most common clinical causes for CP, such as guidewire-induced, balloon catheter bursting, and device oversizing. Moreover, the sealing capacity of covered stent grafts was assessed for each of these models by means of angiography. Results We herein report a rabbit iliac artery perforation model using three different types of vessel perforation that closely mimic the clinical setting, such as guidewire-induced, balloon catheter rupture, and device oversizing. Using the same rabbit iliac perforation model, we additionally assessed the sealing capacity of a covered stent graft for each model. Conclusions The novel rabbit iliac artery perforation models, as described in this report, represent promising animal testing approaches. While their setting is very similar to the real-life context encountered in humans, all three models are based on an animal model that is ideally suited for evaluating the sealing capacity and performance of new medical devices for humans. Keywords: Coronary perforation, Covered stent, Stent design, Guidewire, Balloon catheter, Rupture, Oversizing, Rabbit iliac artery model
Author(s): Alper üner[sup.1,3], Caroline Moerke[sup.1], Anne Wolff[sup.1], Sabine Kischkel[sup.2], Wolfram Schmidt[sup.2], Niels Grabow[sup.2] and Hüseyin Ince[sup.1] Background One of the most common therapeutic procedures in cardiology is percutaneous coronary intervention, [...]