학술논문

Sinogram-based motion correction of PET images using optical motion tracking system and list-mode data acquisition
Document Type
Author Abstract
Source
IEEE Transactions on Nuclear Science. June 2004, Vol. 51 Issue 3, p782, 7 p.
Subject
Nuclear physics -- Research
Language
English
ISSN
0018-9499
Abstract
Motion of the head during brain positron emission tomography (PET) acquisitions has been identified as a source of artifact in the reconstructed image. A number of techniques have been proposed to correct for this motion artifact, but they are unable to correct for a motion during an acquisition. The aim of this study was to develop a sinogram-based motion correction (SBMC) technique to correct directly the head motion during a PET scan using a motion tracking system and list-mode data acquisition. This method uses a rebinning procedure whereby the lines of response (LOR) are geometrically transformed according to the current values of six-dimensional motion data. A Michelogram was recomposed using the rebinned LOR, and the motion-corrected sinogram was generated. In the motion corrected image, the blurring artifact due to the motion was reduced by the SBMC technique. This technique was applied to actual PET data acquired in the list-mode, and demonstrated the potential for real-time motion correction of head movements during a PET acquisition. Index Terms--Motion correction, optical tracking system, PET, sinogram.