Safety and efficacy of flow diverter stents in the treatment of middle cerebral artery aneurysms: a single-center experience and follow-up data
Document Type
Report
Source
Diagnostic and Interventional Radiology. March-April 2023, Vol. 29 Issue 2, p350, 9 p.
Subject
Turkey
Language
English
ISSN
1305-3825
Abstract
Middle cerebral artery (MCA) aneurysms constitute approximately 22% of all cerebral aneurysms. They are generally wide-necked, have accompanying integrated branches, and sometimes have complex anatomy with angulation. Thus, the endovascular [...]
PURPOSE This study aims to evaluate the safety and efficacy of flow diverters (FDs) in the treatment of middle cerebral artery (MCA) aneurysms and share the follow-up (F/U) results. METHODS The treatment and F/U results of 76 MCA aneurysms treated with the flow re-direction endoluminal device (FRED), FRED Jr., and pipeline embolization device (PED) FD stents were evaluated retrospectively. The aneurysm occlusion rates were compared between FDs, and the integrated and jailed branches were evaluated through follow-ups. The oversizing of the stent was compared between occluded/non-occluded aneurysms and integrated branches. RESULTS The mean F/U duration was 32 [+ or -] 6.3 months, and the mean aneurysm diameter was 4.45 mm. A total of 61 (80.3%) aneurysms were wide-necked; 73 (96.1%) were saccular; 52 (68.4%) were located at the M1 segment; and 36 (45.6%) FREDs, 23 (29.1%) FRED Jr.s, and 19 (24.1%) PEDs were used for treatment. The overall occlusion rates for the 6-, 12-, 24-, 36-, and 60-month digital subtraction an-giographies were 43.8%, 63.5%, 73.3%, 85.7%, and 87.5% respectively. The last F/U occlusion rates were 67.6% for FRED, 66.7% for PED, and 60.6% for FRED Jr. (P = 0.863). An integrated branch was covered with an FD during the treatment of 63 (82.8%) aneurysms. A total of six (10%) of the integrated branches were occluded without any symptoms at the last F/U appointment. The median oversizing was 0.45 (0-1.30) for occluded aneurysms, and 0.50 (0-1.40) for non-occluded aneurysms (P = 0.323). The median oversizing was 0.70 (0.45-1.10) in occluded integrated branches and 0.50 (0-1.40) in non-occluded branches (P = 0.131). In-stent stenosis was seen in 22 (30.1%) of the stents at the 6-month F/U and in only 2 (4.7%) at the 24-month F/U. Thus, none of the patients had any neurological deficits because of the in-stent stenosis. Severe in-stent stenosis was seen in two stents. CONCLUSION MCA aneurysms tend to be complex, with integrated branches and potentially wide necks. FD stents are safe and effective in the treatment of MCA aneurysms, and the patency of the side and jailed branches is preserved in most cases. Higher occlusion and lower in-stent stenosis rates are seen with longer F/U durations. KEYWORDS Aneurysm, flow diverter, FRED, MCA, pipeline