학술논문

Adaptive R-wave detection method in dynamic ECG with heavy EMG artifact
Document Type
Conference
Source
2012 IEEE International Conference on Information and Automation Information and Automation (ICIA), 2012 International Conference on. :83-87 Jun, 2012
Subject
Computing and Processing
Communication, Networking and Broadcast Technologies
Electrocardiography
Electromyography
Wavelet transforms
Interference
Noise
Educational institutions
R-wave detection
Dynamic ECG
EMG interference
ECG analysis
Language
Abstract
Recently, the electrocardiogram (ECG) was applied in long-term and dynamic monitoring in daily life, and the R-wave detection is the key in all kinds of ECG monitorings. Electromyographic (EMG) interference is the most common noise in dynamic ECG, and it has great influence on detection of R-wave. Therefore, we propose an adaptive R-wave detection method, which is based on the analysis of EMG interference. In this paper, we first demonstrate different changes of ECG morphology due to different types of EMGs. Then we classify EMG interference by their frequency power ratios, zero-crossing rate and short time energy. The choice of R-wave detection methods (wavelet transform and Empirical Mode Decomposition method) are related to the result of classification. The proposed adaptive R-wave detection method is validated using the data collected in our laboratory, and the experimental results demonstrate that the proposed method is significantly applicable in case of dynamic ECG.