학술논문

Intercomparison of radiation temperature scales between PTB and NMIJ from 1000/spl deg/C to 2500/spl deg/C
Document Type
Conference
Source
Proceedings of the 41st SICE Annual Conference. SICE 2002. SICE 2002 SICE 2002. Proceedings of the 41st SICE Annual Conference. 1:58-62 vol.1 2002
Subject
Robotics and Control Systems
Signal Processing and Analysis
Aerospace
Temperature
Wavelength measurement
Calibration
Lamps
Metrology
Copper
Gold
Silver
Laboratories
Instruments
Language
Abstract
The intercomparison of radiation temperature scales between PTB and NMIJ was carried out in September 2001. PTB sent their radiation thermometer LP3 to Japan and NMIJ made the comparison between the LP3 and NMIJs Topcon radiation thermometers with a comparison blackbody from 1000/spl deg/C to 2500/spl deg/C. Both scales agreed within /spl plusmn/0.8/spl deg/C in the whole range.