학술논문

Single Pixel Performance of a 32 × 32 Ti/Au TES Array With Broadband X-Ray Spectra
Document Type
Periodical
Source
IEEE Transactions on Applied Superconductivity IEEE Trans. Appl. Supercond. Applied Superconductivity, IEEE Transactions on. 31(5):1-5 Aug, 2021
Subject
Fields, Waves and Electromagnetics
Engineered Materials, Dielectrics and Plasmas
X-ray detectors
Energy resolution
Energy measurement
Frequency division multiplexing
Energy calibration
energy resolution
MXS
TES
X-IFU
x-ray detectors
Language
ISSN
1051-8223
1558-2515
2378-7074
Abstract
We are developing a kilo-pixels Ti/Au TES array as a backup option for Athena X-IFU. Here we report on single-pixel performance of a 32 × 32 array operated in a Frequency Division Multiplexing (FDM) readout system, with bias frequencies in the range 1–5 MHz. We have tested the pixels response at several photon energies, by means of a $^{55}$Fe radioactive source (emitting Mn-K$\alpha$ at 5.9 keV) and a Modulated X-ray Source (MXS, providing Cr-K$\alpha$ at 5.4 keV and Cu-K$\alpha$ at 8.0 keV). First, we report the procedure used to perform the detector energy scale calibration, usually achieving a calibration accuracy better than $\sim$0.5 eV in the 5.4–8.9 keV energy range. Then, we present the measured energy resolution at the different energies (best single pixel performance: $\Delta$ E $_{\mathrm{FWHM}}$ = 2.40 $\pm$ 0.09 eV @ 5.4 keV; 2.53 $\pm$ 0.10 eV @ 5.9 keV; 2.78 $\pm$ 0.16 eV @ 8.0 keV), investigating also the performance dependency from the pixel bias frequency and the count rate. Thanks to long background measurements ($\sim$1 d), we finally detected also the Al-K$\alpha$ line at 1.5 keV, generated by fluorescence inside the experimental setup. We analyzed this line to obtain a first assessment of the single-pixel performance also at low energy ($\Delta$ E $_{\mathrm{FWHM}}$ = 1.91 eV $\pm$ 0.21 eV @ 1.5 keV), and to evaluate the linearity of the detector response in a large energy band (1.5–8.9 keV).