Expectation Maximization Reconstruction of PET Image with Non-rigid Motion Compensation
Document Type
Conference
Source
2005 IEEE Engineering in Medicine and Biology 27th Annual Conference Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the. :4453-4456 2005
Subject
Bioengineering
Image reconstruction
Positron emission tomography
Motion compensation
Reconstruction algorithms
Degradation
Time varying systems
Event detection
Motion detection
Histograms
Computational modeling
Language
ISSN
1094-687X
1558-4615
Abstract
Motion artifact degrades PET imaging which often leads to inaccurate quantification of radioactivity concentration. Current motion correction algorithms generally lack the ability to compensate for non-rigid motions as well as perform poorly for frequent non-cyclical motions. In this paper, we derive from first principles a general list mode reconstruction algorithm that accounts for time-varying system responses. We further model the system response function in a way to incorporate the non-rigid motion. The resultant list mode reconstruction algorithm has the ability to compensate for frequent and non-rigid motions, with all detected events utilized. We also provide a histogram reconstruction algorithm that is tailored for gated PET acquisition with reduced usage of computational resource. This algorithm shows good motion compensation capability with computer simulated data and there is no obvious otion artifact visible on the reconstructed image.