학술논문

Mathematical derivation, circuits design and clinical experiments of measuring blood flow volume (BFV) at arteriovenous fistula (AVF) of hemodialysis (HD) patients using a newly-developed photoplethysmography (PPG) sensor
Document Type
Conference
Source
2018 19th International Symposium on Quality Electronic Design (ISQED) Quality Electronic Design (ISQED), 2018 19th International Symposium on. :353-356 Mar, 2018
Subject
Bioengineering
Communication, Networking and Broadcast Technologies
Components, Circuits, Devices and Systems
Engineering Profession
Photonics and Electrooptics
Avalanche photodiodes
Skin
Manganese
Kidney
Handheld computers
Conferences
Technological innovation
Photoplethysmography (PPG) Sensor
Arteriovenous Fistula (AVF)
Blood Flow Volume (BFV)
Language
Abstract
Mathematical derivation of calculating blood flow volume (BFV) at arteriovenous fistula (AVF) using a newly-developed photoplethysmography (PPG) sensor is presented in this work. Also, the readout circuit of the PPG sensor intended to increase the signal-noise ratio (S/NR) is designed and presented in this work. The designed PPG sensor equipped with derived mathematical equations shows high correlation (R 2 = 0.7563) and low error (RMSE = 212 ml/min) as compared to the gold standard.