학술논문

Large-Scale Detector Testing for the GAPS Si(Li) Tracker
Document Type
Periodical
Source
IEEE Transactions on Nuclear Science IEEE Trans. Nucl. Sci. Nuclear Science, IEEE Transactions on. 70(8):2125-2133 Aug, 2023
Subject
Nuclear Engineering
Bioengineering
Detectors
Testing
Preamplifiers
Strips
Flanges
Energy resolution
Silicon
Dark matter
semiconductor device noise
silicon radiation detectors
Language
ISSN
0018-9499
1558-1578
Abstract
Lithium-drifted silicon [Si(Li)] has been used for decades as an ionizing radiation detector in nuclear, particle, and astrophysics experiments, though such detectors have frequently been limited to small sizes ( ${\text {few cm}^{2}}$ ) and cryogenic operating temperatures. The 10-cm-diameter Si(Li) detectors developed for the General Antiparticle Spectrometer (GAPS) balloon-borne dark matter experiment are novel particularly for their requirements of low cost, large sensitive area ( ${\sim 10 \text {m}^{2}}$ for the full 1440-detector array), high temperatures (near −40°C), and energy resolution below 4 keV full-width-at-half-maximum (FWHM) for 20–100-keV x-rays. Previous works have discussed the manufacturing, passivation, and small-scale testing of prototype GAPS Si(Li) detectors. Here, we show for the first time the results from the detailed characterization of over 1100 flight detectors, illustrating the consistent intrinsic low-noise performance of a large sample of GAPS detectors. This work demonstrates the feasibility of large-area and low-cost Si(Li) detector arrays for next-generation astrophysics and nuclear physics applications.