MARIE 2.0: A Perturbation Matrix Based Patient-Specific MRI Field Simulator
Document Type
Periodical
Source
IEEE Transactions on Biomedical Engineering IEEE Trans. Biomed. Eng. Biomedical Engineering, IEEE Transactions on. 70(5):1575-1586 May, 2023
Subject
Bioengineering
Computing and Processing
Components, Circuits, Devices and Systems
Communication, Networking and Broadcast Technologies
Magnetic resonance imaging
Integral equations
Couplings
Computational modeling
Safety
Perturbation methods
Optimization
high-field MRI
magnetic resonance imaging
network parameters
perturbation operator
pseudo-skeleton cross approximation
radio-frequency coils
Language
ISSN
0018-9294
1558-2531
Abstract
High static field MR scanners can produce human tissue images of astounding clarity, but rely on high frequency electromagnetic radiation that generates complicated in-tissue field patterns that are patient-specific and potentially harmful. Many such scanners use parallel transmitters to better control field patterns, but then adjust the transmitters based on general guidelines rather than optimizing for the specific patient, mostly because computing patient-specific fields was presumed far too slow. It was recently demonstrated that the combination of fast low-resolution tissue mapping and fast voxel-based field simulation can be used to perform a patient-specific MR safety check in minutes. However, the field simulation required several of those minutes, making it too slow to perform the dozens of simulations that would be needed for patient-specific optimization. In this paper we describe a compressed-perturbation-matrix technique that nearly eliminates the computational cost of including complex coils (or coils and shields) in voxel-based field simulation of tissue, thereby reducing simulation time from minutes to seconds. The approach is demonstrated on a wide variety of head+coil and head+coil+shield configurations, using the implementation in MARIE 2.0, the latest version of the open-source MR field simulator MARIE.