Abstract
A preclinical insert for small animal simultaneous SPECT and MR imaging, in particular for imaging mouse brains, is presented. It consists of ten static magnetic resonance imaging (MRI)-compatible gamma cameras based on tiles of silicon photomultipliers readout by a multichannel ASIC and coupled to 5 cm $\boldsymbol \times $ 5 cm CsI(Tl) scintillators and to an MRI-compatible multipinhole collimator. Calibration and image reconstruction algorithm are illustrated. Mutual compatibility is demonstrated along with imaging performance that is comparable with other non-MR micro-SPECT systems: 0.9 mm tomographic spatial resolution across a transverse field of view of 15.6 mm, 12% energy resolution (at 140 keV), and 1105 cps/MBq sensitivity. Experimental results with phantoms (glass capillaries of 290 $ \mu \text{m}$ diameter and a mini Derenzo) are presented.